Table 1. Accuracy, precision, recovery, and detection and quantitative limits of nucleotides using high-performance liquid chromatography with diode array detector (HPLC-DAD)
Nucleotides | Spiked amount (mg/kg) | Intra-day | Inter-day | LOD (mg/L) | LOQ (mg/L) | Coefficient correlation (r2) | Recovery2)(%) |
Accuracy1)(%) | Precision (CV, %) | Accuracy (%) | Precision (CV, %) |
CMP | 2 | 101.06±1.43 | 1.51 | 102.72±1.71 | 3.33 | 0.09 | 0.31 | 0.9999 | 101.06±1.43 |
UMP | 2 | 84.90±2.27 | 3.27 | 84.69±1.06 | 6.82 | 0.11 | 0.37 | 0.9999 | 84.90±2.27 |
AMP | 2 | 96.18±3.59 | 3.71 | 98.01±1.84 | 2.33 | 0.12 | 0.38 | 0.9999 | 96.18±3.59 |
GMP | 2 | 96.62±3.64 | 1.56 | 99.26±2.32 | 2.79 | 0.11 | 0.39 | 0.9999 | 96.62±3.64 |
IMP | 2 | 102.28±1.52 | 5.21 | 99.33±2.89 | 4.20 | 0.14 | 0.45 | 0.9999 | 102.28±1.52 |
Accuracy=Measured concentration/Spiked concentration (Mean).
Mean±SD (n=3).
HPLC-DAD, high-performance liquid chromatography with diode array detector; CV, SD of measured concentration/means of measured concentration; LOD, limit of detection, signal-to-noise (S/N) ratio=3; LOQ, limit of quantitation, S/N ratio=10; CMP, cytidine 5′-monophosphate; UMP, uridine 5′-monophosphate; AMP, adenosine 5′-monophosphate; GMP, guanosine 5′-monophosphate; IMP, inosine 5′-monophosphate.